| Catalog Number |
MZ9281 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.The date of event is unknown; bd awareness date used.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd maxzero extension set was damaged.The following information was provided by the initial reporter: our pediatric units recently implemented mz9281, and we have now received 5 reports of the luer collar cracking.All were lot 25085808.Please mention the affected number of patients with respect to the event dates.4 describe any patient¿s harm, injury, complication or negative outcome that occurred as a result of the event.Chemotherapy exposure.
|
| |
|
Event or Problem Description
|
|
Sample investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Description of event
it was reported by a customer that the luer collar was cracking.Investigation results
the customer returned 1 sample of trifuse material mz9281, lot 25085808.The sample was returned under (b)(4).Upon initial visual examination, the set verified the report of component damage, and the male luer spin collar was cracked and broken.The samples were sent for further investigation to the supplier of the component.Device / batch history records (dhr/bhr) for batch 25085808 were reviewed to determine if there was a potential manufacturing or material issue related to the reported complaint.Review of in-process and final inspection records confirmed that all required inspections were completed and met established acceptance criteria.The supplier of the component confirmed the failure, but was unable to define the root cause or trace it to their process.Due to the ongoing issue with the mz9283 quadfuse and mz9281 trifuse sets, the investigation has been escalated to a cross-functional team, and a funded project has been opened.The project will further investigate the issue.
|
| |
|