| Model Number |
132-20-50-250 |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The implant remains in situ.Radiographic images were provided which confirm the event of the screw passing through the spacer.A review of the device history records could not be performed as the identifying lot number was not provided.Based on the information provided, the root cause could not be determined.Alphatec spine is submitting this report to comply with the fda regulations 21 cfr 803.Alphatec spine has made reasonable efforts to provide as much relevant information as available.Any field that is left blank was not known at the time of this submission.If additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
A patient underwent an anterior lumbar interbody fusion procedure on (b)(6) 2026.It was reported that during the procedure the screw advanced through the spacer, bypassing the locking mechanism.
|
| |
|
Additional Manufacturer Narrative
|
|
E4.Yes.
|
| |
|
Search Alerts/Recalls
|