|
The monopolar curved scissors (mcs) instrument has been returned and evaluated by the failure analysis (fa) team.The instrument was found to have a broken tube adapter.For clarification, the tip accessory cannot be installed on the instrument due to a broken tube adapter.This component is located at the distal end of the instrument and serves as the interface between the instrument and the user installed tip accessory.The broken piece, measuring approximately 1.7mm x 4.2mm was not returned with the instrument.Visual inspection was performed and there was no damage to the snake wrist cables, housing and shaft.Fa found additional observations that are related to the customer reported complaint: the instrument was found to have a detached fragment.Visual inspection was performed and the detached fragment was broken from the tube adapter.The area of the detached fragment measures approximately 1.7mm x 4.2mm size and located at the distal tip.The instrument was found to have scratch marks and abrasions on the tube adapter.During visual inspection, several scratch marks and abrasions was observed on the tube adapter, located at the distal tip of the instrument.The complaint was confirmed by failure analysis.The probable root cause of a damaged tube adapter is attributed to either tip accessory installation issues or damage during use.
|