Catalog Number 5531-G-309-E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was provided: revised a right triathlon cementless knee due to infection.The patient had an insert exchange by on (b)(6) 2025.Dr.Cemented a ts femur with an all-poly tibia as a 1st stage revision.
 
Additional Manufacturer Narrative
The following devices were also listed in this report: cat#;device name; lot#: 5552-l-299;tritanium patella-asymmetric;vk1n1.5536b300;tritanium bplate triathlon s3;ctd117999.5517f302;triathlon p/a cr beaded #3r;x6uhu.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding infection involving a triathlon insert was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the lot referenced.There has been 1 other similar event for the sterile lot referenced.Also relates to infection.A review of both the sterilization data and microbiology data was performed for this sterile lot commonality.It was identified that the sterilization data was within specification and no microbiology excursions were noted for this lot and as such no further review is required.Conclusions: it was reported that the patient was revised due to infection.All stryker products are sold as sterile either by being validated to a minimum sterility assurance level sal of 10^-6 or aseptically filled under a validated process in accordance with applicable external standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT NO 3 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
alessandra chavez
3365 enterprise ave
weston, FL 33331
9546280700
MDR Report Key25315023
Report Number0002249697-2026-00475
Device Sequence Number5295040
Product Code JWH
UDI-Device Identifier07613327336955
UDI-Public07613327336955
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K242831
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5531-G-309-E
Device Lot Number5K4T37
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexUnknown
-
-