MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC / NASOINTESTINAL FEEDING TUBE WITH STYLET; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC / NASOINTESTINAL FEEDING TUBE WITH STYLET; DH CPK NG TUBES Back to Search Results
Model Number 20-9551
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.No sample was returned for evaluation.However, customer-provided video documentation confirmed improper connection/misconnection between the ng tube y-port, connector extension set, and administration set.Investigation also identified that a non-avanos transition connector appeared to have been used with the device.The device history record for lot: 30127679 was reviewed and the product was produced according to product specifications.Root cause could not be determined.All information reasonably known as of 27 may 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
The customer reported that a nasogastric tube could not be properly connected during use in the icu.The condition was reportedly identified after approximately six days of use.No patient injury or medical treatment was reported; however, the tube was replaced with a new tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC / NASOINTESTINAL FEEDING TUBE WITH STYLET
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key25315421
Report Number9611594-2026-00369
Device Sequence Number14794454
Product Code KNT
UDI-Device Identifier00350770458137
UDI-Public(01)00350770458137(240)104581302(17)260414(10)30127679
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K220588
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/14/2026
Device Model Number20-9551
Device Catalogue Number20-9551
Device Lot Number30127679
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2023
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-