• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VANTIVE US HEALTHCARE LLC PRISMAFLEX SET; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107640US
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Prismaflex st150 set has been temporarily approved for use in the us under emergency use authorization (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the luer connector of a prismaflex st 150 set was observed to be broken during continuous renal replacement therapy.Treatment was discontinued when "rising filter pressures" occurred and on disconnection, the access connector was noted to be stuck inside the patient catheter.The extracorporeal blood was not returned.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed that the cone of the set return male luer lock was broken.This component is provided preassembled by a vantive supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the broken luer connector was determined to be related to supplier product design and manufacturing.Corrective action preventive action and change control records were previously opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
deerfield IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - MEYZIEU
7, av. lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
510 lake cook rd
deerfield, il, IL 60015
8153072966
MDR Report Key25316158
Report Number8010182-2026-00428
Device Sequence Number14775230
Product Code KDI
UDI-Device Identifier00085412917696
UDI-Public0085412917696
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K212216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number107640US
Device Lot Number25G0025CB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2026
Supplement Date Received by Manufacturer07/01/2026
Initial Report FDA Received Date05/28/2026
Supplement Report FDA Received Date07/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
8010182-12/17/25-001-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRISMAX MACHINE.
Patient SexMale
-
-