| Catalog Number |
107640US |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Prismaflex st150 set has been temporarily approved for use in the us under emergency use authorization (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the luer connector of a prismaflex st 150 set was observed to be broken during continuous renal replacement therapy.Treatment was discontinued when "rising filter pressures" occurred and on disconnection, the access connector was noted to be stuck inside the patient catheter.The extracorporeal blood was not returned.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed that the cone of the set return male luer lock was broken.This component is provided preassembled by a vantive supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the broken luer connector was determined to be related to supplier product design and manufacturing.Corrective action preventive action and change control records were previously opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|