| Model Number |
LIFEVEST WCD 5100 SYSTEM |
| Medical Device Problem Code |
Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation of battery charger has been completed.The reported problem (will not power on) has been confirmed.Upon investigation the battery charger was unable to power on.The cause for the failure was isolated to a defective pcb.The root cause for the defective pcb could not be positively identified.There was no adverse event that resulted from the damaged charger.
|
| |
|
Event or Problem Description
|
|
A us distributor reported that a battery charger will not power on.
|
| |
|