• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SPEEDPLATE; STAPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SPEEDPLATE; STAPLE Back to Search Results
Model Number SK50
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Swelling/ Edema (4577)
Date of Event 04/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed on (b)(6) 2026 due to swelling.Upon removal, there were no deficiencies or malfunctions found with any tmc device and the patient's bones were completely fused.No other patient outcomes were reported as a result of this event.
 
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed on (b)(6) 2026 due to swelling.Upon removal, there were no deficiencies or malfunctions found with any tmc device and the patient's bones were completely fused.No other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records were reviewed, and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.The most likely cause of the reported event could not be determined based on the available information and without the return of the device.There were no deficiencies or malfunctions alleged/found with any tmc device, nor were there any reported issues with placement of the device or healing of the surgical site.The company will supplement this mdr as necessary and appropriate.Additional tmc device explanted in the same revision surgery was reported in 3011623994-2026-00129, 3011623994-2026-00130, 3011623994-2026-00131.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY SPEEDPLATE
Common Device Name
STAPLE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
artillia walker
100 palmetto park place
ponte vedra, FL 32081
9049131020
MDR Report Key25317961
Report Number3011623994-2026-00132
Device Sequence Number5368612
Product Code JDR
UDI-Device Identifier00810111221788
UDI-Public(01)00810111221788(17)300622(10)134066
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K251135
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSK50
Device Lot Number134066
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2026
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-