|
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE "PLASMA-OVALBUTTON", BUTTON, OVAL, 24 FR., 12°-30°
|
Back to Search Results |
|
| Model Number |
WA22766S |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of this complaint is traced to cause traced to component failure expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the hf-resection electrode to the service center for a separate issue.During the device analysis, the investigation indicates that the electrode was bent near the middle section.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|
|
|