(b)(4).Date sent: 5/28/2026.D4: batch #unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: any pictures or video available? an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore, it has been determined that no corrective and preventive action is required at this time.However, if the product is received at a later date, the investigation will be updated as applicable.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during a thoracoscopic lung resection procedure, ethicon 4000 with a green cartridge was used on the left lower bronchus.When checking the staple line, six malformed staples (approximately 80-90%) were observed at the 2d staple line on the main bronchus side of the stump.The staples formed in a u-shape and were considered not to have contacted the anvil.Additional suturing and reinforcement with mediastinal fat tissue were performed, and the procedure was completed.The patient condition after surgery was stable.The cartridge color was green.There were no adverse consequences to the patient.No additional information was provided.
|