Model Number EVOLUTR-26-US
Medical Device Problem Codes Degraded (1153); Activation Failure (3270); Patient Device Interaction Problem (4001); Thickening of Material (4056)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 05/26/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that approximately 8 years and 8 months following the implant of a transcatheter aortic valve valve-in-valve in an indwelling surgical aortic valve, the valve failed due to under-expansion and thickened leaflets with pannus and thrombus formation.
 
Event or Problem Description
It was reported, this transcatheter aortic valve (tav) was implanted within a failed 25mm medtronic (mosaic) surgical aortic valve.Approximately eight years, eight months following the tav implant, the tav failed due to under-expansion and thickened leaflets with pannus and thrombus formation.Additional information was received to report intent for the patient to undergo reintervention for the degenerated tav.
 
Additional Manufacturer Narrative
Updated data: b1.Adverse event or product b2.Outcome attributed to adverse event b5.A second paragraph was added.H1.Type of reportable event h6.Device code additional codes.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT R TRANSCATHETER AORTIC VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key25318921
Report Number9617601-2026-03457
Device Sequence Number20140988
Product Code NPT
UDI-Device Identifier00643169637993
UDI-Public(01)00643169637993(17)190530
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2019
Device Model NumberEVOLUTR-26-US
Device Catalogue NumberEVOLUTR-26-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2026
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date05/28/2026
Supplement Report FDA Received Date06/23/2026
Date Device Manufactured06/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-