| Model Number |
G148 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device remains in service and was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported that the health care professional (hcp) called technical services (ts) for a review of reports for this cardiac resynchronization therapy defibrillator (crt-d).Upon review, ts noted this crt-d exhibited far-field oversensing on the right atrial (ra) channel.No adverse patient effects were reported.This crt-d remains in service.
|
| |
|