It was reported that during use, the cuff pressure begun to drop that was monitored via the pressure gauge.Attempted to reinflate the cuff, however, the pressure could not be maintained within the normal range of 20¿30 cmh2o.The tube was subsequently replaced.As the case was nearly completed, the second tube was removed after approximately three hours of use, even though the cuff pressure was again found to have dropped below 20 cmh2o, falling outside the acceptable range.There was no patient injury.For pli 30: medtronic's initial evaluation of the incident device found that the cuff was torn pli 40: unauthorized rerurn.
|
|
D10 concomitant products: 87480, 87480 shiley oral nasal intermediate 8.0, (lot# 25h0257jzx) 87480, 87480 shiley oral nasal intermediate 8.0, (lot# 25h0257jzx) 87480, 87480 shiley oral nasal intermediate 8.0, (lot# 25h0257jzx) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the sample was received outside its pouch and it was observed the cuff was torn.An inflation/deflation test was performed by applying 25 cubic centimeters of air to the cuff.The cuff deflated immediately.It was reported that the cuff pressure begun to drop that was monitored via the pressure gauge.Attempted to reinflate the cuff, however, the pressure could not be maintained within the normal range of 20¿30 cmh2o.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not overinflate the cuff.It is recommended that the cuff pressure should not exceed 25cmh2o.Overinflation can result in tracheal damage, rupture of the cuff with subsequent deflation, or in cuff distortion which may lead to airway blockage.Each tube¿s cuff, pilot balloon and valve should be tested by inflation prior to use.If dysfunction is detected in any part of the inflation system, the tube should not be used and should be returned to the manufacturer.Various bony anatomical structures (e.G., teeth, turbinates) within the airway or any intubation tools with sharp surfaces might jeopardize cuff integrity.Damage to the thin-walled cuff during insertion will subject the patient to the risk of extubation and reintubation.If the cuff is damaged, the tube should not be used.Deflate the cuff prior to repositioning the tube.Movement of the tube with the cuff inflated could result in patient injury, req uiring possible medical intervention or damage to the cuff.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|