BECKMAN COULTER, INC. DXC 700 AU CHEMISTRY ANALYZER; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
|
Back to Search Results |
|
| Model Number |
DXC 700 AU-10E CHEMISTRY ANALYZER DXC 700 AU WITH ISE |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Beckman coulter field service engineer (fse) evaluated the dxc 700 au chemistry analyzer and identified the probable cause as defective buffer valves.The fse replaced the buffer valves to resolve the issue.No further issues were reported and the customer was satisfied.Section a2, a3, a4, a5 and a6: information not provided by customer.The beckman coulter internal identifier is case (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported the generation of erroneously high ion-selective electrolytes (ise) patient results, including sodium (na), chloride (cl) and potassium (k), on their dxc 700 au clinical chemistry analyzer.There was no report of change to treatment, injury, or death associated with event.The customer did not provide patient data or demographics.
|
| |
|