| Catalog Number |
50796120 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint could be confirmed, due to feedback from other complaints.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.Inspection of the received information/evidences are done or will be done, in the proceedings of the nc.Based on investigation, the root cause was attributed to a design related issue.
|
| |
|
Event or Problem Description
|
|
It was reported that the replacement delivery of the carbon rods were received.They were processed once in the reprocessing/sterilization unit for medizinprodukte (aemp (aufbereitungseinheit for medizinprodukte)), unfortunately the carbon rods broke again.
|
| |
|
Search Alerts/Recalls
|