| Model Number |
G46206 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
It was reported that upon opening the packaging of an ncircle tipless stone extractor, the wire of the extractor was observed to be broken prior to use.The user replaced the device with a new similar-like device to complete the ureteroscopic lithotripsy (ursl) procedure without difficulties.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
|
| |
|
Search Alerts/Recalls
|