| Model Number |
BF-XT190 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: component failure from environmental temperature problem.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The bronchovideoscope was returned to the service center with a report of minor leak.This does not indicate an mdr reportable event.However, an mdr reportable failure was found during testing at the service center.During inspection of the device, burnt c-cover was found.There was no patient involvement.
|
| |
|