ETHICON INC. PROLENE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8523H
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).Date sent to the fda: 5/28/2026.H6: component code: g07002 - device not returned.H6: component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Additional information was requested, the following was obtained: 1.Were there any adverse events (i.E.Infection/complication etc), patient consequences or harm to the patient? ans: no.2.Was the surgery completed successfully? if yes, how? ans: yes, open new sutures.3.Please provide the product code and lot number for the impacted product.They were not stated in the complaint.Ans: product code: 8523h.Lot number: ubbclh.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photo shows a winding former with the suture dispensed inside a plastic bag.The photo is not clear because of the bag.As the device was not returned, an analysis investigation could not be performed.Based on the photo review, the event reported is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.We value your assistance in providing us an opportunity to evaluate the reported event as all information reported to our company is critical to our continuous improvement efforts.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported that a patient underwent a cabg procedure on an unknown date and suture was used.It was reported by sister that a sutures bifurcated during cabg procedure.The issue was identified during intraoperative use.There were no patient consequences reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
*   00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25322583
Report Number2210968-2026-05894
Device Sequence Number5181310
Product Code GAW
UDI-Device Identifier10705031016590
UDI-Public(01)10705031016590(17)290131(10)UBBCLH
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8523H
Device Lot NumberUBBCLH
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2026
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-