COOK INC NCIRCLE TIPLESS STONE EXTRACTOR; FFL, DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number G46206
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
G4 ¿ pma/510(k) #: exempt.H3 - device evaluation anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that the ncircle tipless stone extractor could be opened and closed approximately twice, but after that it became unusable.The procedure was completed using another device of the same specifications.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Corrected information: h6 adverse event problem (annex a and g).Investigation ¿ evaluation.It was reported that the ncircle tipless stone extractor could be opened and closed approximately twice, but after that it became unusable.The procedure was completed using another device of the same specifications.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control procedures, specifications and instructions for use (ifu), as well as a visual inspection of the returned device, were conducted during the investigation.One used complaint device was returned without packaging or label.The support sheath is severed 1 mm past handle.The basket sheath is severed also at the same location as the support sheath.The cannulated handle is kinked over at the nose of the male luer lock adapter.Cook has concluded that the device was manufactured to specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot and the related subassembly lots did not reveal any related quality control discrepancies relevant to the failure mode.A complaint history search identified no other events reported for a related failure mode.Based on the available information, cook has concluded that there is no evidence suggesting product manufactured out of specification exists either in house or in the field.Cook also reviewed product labeling.The ifu packaged with the device contains the following in relation to the reported failure mode: the product ifu, t_ntse_rev1 was reviewed and includes the following: precautions ¿ do not use excessive force to manipulate this device.Damage to the device may occur.Based on the information provided, inspection of the returned device, and the results of the investigation, a potential root cause for this event could not be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute toa death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE TIPLESS STONE EXTRACTOR
Common Device Name
FFL, DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key25323386
Report Number1820334-2026-00568
Device Sequence Number14776054
Product Code FFL
UDI-Device Identifier00827002462069
UDI-Public(01)00827002462069(17)290317(10)17038875
Combination Product (Y/N)N
Initial Reporter CountryJP
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG46206
Device Catalogue NumberNTSE-015115
Device Lot Number17038875
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2026
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-