E.3.Other occupation: product marketing manager there were multiple medical device types reported to be involved.The information for the additional device type is as follows.D.2.B medical device type: jka.D.2.A common device name: tubes, vials, systems, serum separators, blood collection.There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k213670; k231373.Investigation summary: bd received 4 photos for investigation.Evaluation of the photos confirmed the customer¿s indicated failure mode.Bd also inspected 100 retained samples from production lots, with no visible defects observed.Furthermore, functional tests were performed on 10 retained samples from each lot, and all tubes were found to be within specification limits.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: draw volume.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for the identification of emerging trends.
|