Catalog Number DSF1433
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 05/15/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2026, a patient underwent an endovascular procedure to treat an abdominal aortic aneurysm using a gore® excluder® conformable aaa endoprosthesis device, gore® excluder® aaa endoprosthesis system devices, and gore® dryseal flex introducer sheaths (16fr and 14fr).A 16fr sheath was inserted into the left external iliac artery (leia), and a trunk-ipsilateral leg was inserted through the leia.During deployment of the trunk-ipsilateral leg, it migrated distally; it was fully deployed in that position.The length of the migration was unknown.An aortic extender was deployed to extend the trunk-ipsilateral leg proximally.A 14fr sheath was inserted into the right external iliac artery (reia), and a contralateral leg device was inserted through the reia.Vascular access angiography revealed dissections in the right and left external iliac arteries.Stents were placed in the right and left external iliac arteries to treat the dissections.Reportedly, the minimum diameter of the patient's leia was 3.0 mm; it was smaller than the outer diameter of the 16fr sheath (6.0 mm).Also, the minimum diameter of the patient's reia was 2.6 mm; it was smaller than the outer diameter of the 14fr sheath (5.3 mm).The physician considered that the access vessels degenerated due to the patient's paraplegia.The patient tolerated the procedure.
 
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d1101: this complaint includes information indicating the reported complications relate to use of a sheath size that was too large for the patient access vessel anatomy per the device instructions for use.The device instructions for use provide warning and instruction to ensure the vessel diameter is adequate to accommodate the device.Please note that the same patient was captured in medwatch report number 3007284313-2026-04814.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
mana yamamoto
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key25326085
Report Number3007284313-2026-04815
Device Sequence Number13250358
Product Code DYB
UDI-Device Identifier00733132630004
UDI-Public(01)00733132630004(17)270902
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberDSF1433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2026
Initial Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexMale
-
-