• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX¿; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX¿; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AS075
Medical Device Problem Codes Detachment of Device or Device Component (2907); No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a repair request and no alleged product deficiencies were reported.No patient impact reported.Additional information was received stating that detached bearings were found during evaluation.
 
Additional Manufacturer Narrative
H3: product analysis # (b)(4): evaluation determined that the bearing internal tube is detached.The likely cause of failure is due to inadequate preventative maintenance.It was also noted the spacers are corroded.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX¿
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25327616
Report Number1625507-2026-01536
Device Sequence Number9676147
Product Code HBB
UDI-Device Identifier00643169087583
UDI-Public(01)00643169087583(11)161206
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAS075
Device Catalogue NumberAS075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2024
Initial Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-