| Model Number |
MR8-AF01 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis :evaluation determined that the bearing were fell out issue was confirmed.The likely cause of the failure is due to internal tube corrosion.It was also noted that distal bearing detachment.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the bearing were fell out during a procedure.It was reported that there was no patient impact reported. additional information was received stating that distal bearing was detached were found during evaluation.
|
| |
|