|
The lens was returned adhered to the non-serialized side of the lens case lid (lens case base not returned).Blood and solution was dried on the lens.The optic was cut into two portions, typical of insertion and removal.Both cut optic portions had scratches and cracks on the anterior and posterior sides of the lens near the cut damaged area.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The file indicated the use of a qualified cartridge and handpiece.The associated viscoelastic was not provided.It is unknown if a qualified product was used.The reported broken haptic was observed.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of blood, surgical solution, and the condition of the returned sample, the damage is most likely related to customer handling.The associated viscoelastic was not provided.It is unknown if a qualified product was used.The instructions for use (ifu) instructs: company foldable intra ocular lens (iol)s are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Use the company cartridge at operating room temperatures between 18 degree c (64 degree f) and 23 degree c (73 degree f).Temperature greater than 23 degree c (73 degree f).If the operating room temperature is too high (> 23 degree c / 73 degree f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|