MAUDE Adverse Event Report: TORNIER INC TORNIER PERFORM REVERSED STANDARD GLENOSPHERE DIA 42MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC TORNIER PERFORM REVERSED STANDARD GLENOSPHERE DIA 42MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWJ014
Medical Device Problem Codes Loss of Osseointegration (2408); Device Dislodged or Dislocated (2923); Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Inadequate Osseointegration (2646)
Date of Event 05/06/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported that the patient underwent a right primary reverse shoulder arthroplasty.At approximately two-year follow-up, the patient presented with shoulder pain and reduced range of motion.The surgeon noted reported non-compliance with postoperative rehabilitation and a history of potential trauma while climbing in and out of a truck.During revision surgery, the glenoid component was found to be loose and dissociated from the glenoid bone and was easily removed.The humeral component remained well fixed and was retained.The polyethylene insert exhibited eccentric wear, and grey synovial tissue was observed and excised.Copious lavage was performed.New glenoid components were implanted with the addition of synthetic bone graft.A new humeral polyethylene insert was placed.The shoulder was reduced and found to be stable through range of motion.
 
Event or Problem Description
It was reported that the patient underwent a right primary reverse shoulder arthroplasty.At approximately two-year follow-up, the patient presented with shoulder pain and reduced range of motion.The surgeon noted reported non-compliance with postoperative rehabilitation and a history of potential trauma while climbing in and out of a truck.During revision surgery, the glenoid component was found to be loose and dissociated from the glenoid bone and was easily removed.The humeral component remained well fixed and was retained.The polyethylene insert exhibited eccentric wear, and grey synovial tissue was observed and excised.Copious lavage was performed.New glenoid components were implanted with the addition of synthetic bone graft.A new humeral polyethylene insert was placed.The shoulder was reduced and found to be stable through range of motion.
 
Additional Manufacturer Narrative
This device was concomitant and did not contribute to the reported failure.The complaint is closed without further investigation.If additional information becomes available, the record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED STANDARD GLENOSPHERE DIA 42MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
Tornier Inc
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key25331633
Report Number0001649390-2026-00502
Device Sequence Number18549030
Product Code KWS
UDI-Device Identifier00846832092109
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWJ014
Device Lot NumberAK3038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2026
Supplement Date Received by Manufacturer08/19/2026
Initial Report FDA Received Date05/29/2026
Supplement Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
-
-