• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H180J
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.The device evaluation found the air/water channel had white foreign material the root cause of the white foreign material could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the use of products that contain silicone can cause deposits of white foreign material.Silicone is for example included in simethicone, a defoaming agent, lubricants and so on.If the customer used them, there was risk that foreign material remained in the channel even if the reprocessing was conducted in accordance with the instructions for use (ifu).Simethicone and petroleum/oil/silicone-based lubricants are non-water soluble and thus not recommended for use by olympus.The event can be detected/prevented by following the instructions for use which state: ¿nothing other than sterile water should be used for air/water feeding.No additives should be put into the sterile water.Non-sterile water may cause patient cross-contamination and/or infection.¿ the date of the event is unknown.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
 
Event or Problem Description
The customer returned the gastrointestinal videoscope to the service center for a separate issue.During the analysis, it was observed that the air/water channel had white foreign material.The root cause of the white foreign material could not be identified.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25334033
Report Number9610595-2026-43853
Device Sequence Number9767880
Product Code FDS
UDI-Device Identifier04953170339110
UDI-Public04953170339110
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-H180J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2026
Initial Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-