BOSTON SCIENTIFIC CORPORATION CONCEAL; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 720185-01
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Model number/catalog number: 72404232-10, serial number: n/a, batch/lot number: 1100824198, model/catalog description: cx preconnect ms 18cm ps 10cm tl iz, unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.Device migration is acknowledged within the directions for use (dfu) and are not related to off-label use or failure to follow instructions.A risk review was completed, and migration is defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.The reported migration is a known risk associated with this device type and indicated as such in the instructions for use.Based on the information available, the conclusion code of known inherent risk of device was assigned to this investigation.
 
Event or Problem Description
It was reported that the reservoir of this inflatable penile prosthesis (ipp) had migrated out of the space of retzius to the scrotal area.A surgical procedure was performed to remove and replace the reservoir.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCEAL
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25334399
Report Number2124215-2026-28933
Device Sequence Number9861084
Product Code JCW
UDI-Device Identifier00878953005669
UDI-Public(01)00878953005669(17)270827(10)1100785859
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number720185-01
Device Catalogue Number720185-01
Device Lot Number1100785859
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2026
Initial Report FDA Received Date05/29/2026
Date Device Manufactured08/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
-
-