|
The device was returned to olympus for inspection.A root cause could not be identified.Based on past investigation results, the reported events are inferred to have occurred through the following mechanism: 1.Grasping section was closed without grasping anything while the output in seal & cut mode was activated, (this includes after tissue resection) causing the tissue pad to wear away.2.The non-insulated area of the grasping section and the probe came into contact due to wear of the tissue pad.3.The output in seal & cut was activated while the non-insulated area of the grasping section was in contact with the probe.As a result, a contact mark was developed, and it was causing a probe damage error.The event can be prevented by following the instructions for use which state: should any irregularity (error window, abnormal noise, abnormal output, abnormal operation, abnormal appearance, etc.) or malfunction be observed while using the thunderbeat, stop the use and withdraw the instruments from the body cavity.Do not withdraw the transducer plug from the ultrasonic generator.Check the equipment by referring to chapter 8, ¿troubleshooting¿, in the instruction manual for the ultrasonic generator.If the irregularity remains after troubleshooting, replace the transducer.If this does not resolve the issue, replace the thunderbeat instrument.If the irregularity remains unresolved, contact olympus.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|