| Model Number |
4088 |
| Medical Device Problem Codes |
Over-Sensing (1438); Failure to Read Input Signal (1581); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device remains in service; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported that this pacemaker system exhibited t wave oversensing and one undersensed r wave in the right ventricular (rv) channel.Additionally, it was noted that the r waves appeared to be variable.Technical services (ts) was contacted and recommended possible reprogramming options.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|