| Model Number |
M301 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This icm was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Neither visual examination nor x-ray examination of the device identified anomalies.The device case was removed to facilitate inspection and testing of the internal components.The battery was removed and replaced with an external power source.The battery was forwarded for detailed analysis.This analysis identified a latent current leakage path within the battery that resulted in the observed premature battery depletion.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of "cause traced to device design".
|
| |
|
Event or Problem Description
|
|
It was reported that there were no field allegations made against the performance of this insertable cardiac monitor (icm) device.The device was explanted after being implanted and in-service for approximately 2 years.The device was returned, and device analysis was completed.No adverse patient effects were reported.
|
| |
|