The device was returned to olympus for inspection when the reportable malfunction was observed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed issue was attributed to wear problems, however, a definitive root cause could not be established.It is likely the following led to the malfunction: 1.During output activation in seal & cut mode, nothing was grasped between the grasping section and the distal end of the probe (including after tissue resection).As a result, the tissue pad was worn out.2.Due to friction between the grasping section and the distal end of the probe, abnormal heat was generated, which may have caused the tissue pad to partially peel off.The event can be detected/prevented by following the applicable chapters in the instructions for use which state: ¿ do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ if any irregularity is observed, take proper remedial actions by referring to chapter 8, ¿troubleshooting¿ in the instruction manual for the ultrasonic generator.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|