| Model Number |
D141 |
| Medical Device Problem Codes |
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a clinic visit, this implantable cardioverter defibrillator (icd) was found to have recorded a code 1005 which indicates an open circuit condition was detected during shock delivery.The health care professional (hcp) called technical services (ts) for a consult review of the stored electrogram (egm).Upon review, the egm showed that patient was in a true ventricular tachycardia (vt) arrhythmia.The device delivered a shock which slowed the rhythm rate to fall below rate cutoffs and patient rhythm self-terminated.During the delivery of the shock high out of range shock impedances were detected on the right ventricular (rv) lead which led to the code 1005 to be recorded.Ts recommended a replacement of the lead system.At this time, no additional patient effects were reported.At this time, the icd and rv lead remain in service.
|
| |
|