| Model Number |
UHI-4 |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.In addition, the investigation found that the unit fails secondary piping leakage test, but a substitute test emergency pressure (ep) valve fixed the issue.Based on the results of the investigation, the most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the evaluation, the high flow insufflation unit fails secondary piping leakage test, but a substitute test emergency pressure (ep) valve fixed the issue.There were no reports of patient involvement.
|
| |
|