| Model Number |
CF-H180AL |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction found during device evaluation is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the colonovideoscope which is an olympus asset with no reported complaint.During the device analysis, corroded nozzle was found.It was determined that the nozzle needed to be replaced.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|