| Catalog Number |
32-483820 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tasp was returned fractured.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5; d4; g3; h2; h4; h6
- visual inspection of the returned device exhibits signs of repeated use and part of the insertion and dovetail feature has fractured off.All pieces were not returned.Measurements taken were conforming.-review of device history records identified no related deviations or anomalies during manufacturing.- this device is used for treatment.- reported event is not related to a combination of products; therefore a compatibility review is not applicable.-review of complaint history identified additional similar complaints for the reported items and no additional complaints for part and lot combinations (for all time).-no medical records were provided.- the lot has been in the field for approximately 6+ years.It is unknown for how many times the device is being used.The root cause can be contributed to normal wear and tear of device.-no corrective actions, preventive actions, or field actions resulted after investigation of this event.Complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|