| Model Number |
436909 |
| Medical Device Problem Codes |
Failure to Capture (1081); Ambient Noise Problem (2877); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Loss of capture events seen on the rv channel.Small deflections could be seen upon review suggesting capture except for on one paced event; however, no intrinsic event conducted, so the assessment is isoelectric capture.Noise could be seen on the farfield, atrial and ventricular channels simultaneously, suggesting possible emi.However there was significant atrial undersensing as well as other recordings with pixelated farfield signal, possibly suggesting lead integrity issue.Lead currently remains implanted.No adverse patient events were reported.
|
| |
|