• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECH60C
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a segmentectomy procedure, the stapler was deployed and the knife returned with no problem at the fifth firing.However, although the closing trigger was released and grasping force was loose, the anvil jaw remained closing.The anvil jaw was opened when using the forceps and pulled.After the sales rep tried closing and opening the defective device, although the lever remained open, it could feel for sure that the anvil got caught on something and kept closed.The cartridge color was green.Another device was used to complete the case.There were no adverse consequences to the patient.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 6/01/2026.D4: batch #a98r9p.Additional information was requested, and the following was obtained: was the device locked on tissue with the jaws closed?=>yes if yes, how was the device removed?=>yes.What troubleshooting steps were attempted to open the device (e.G.Squeezing the closing trigger while simultaneously depressing the anvil release button, knife reverse switch, manual override)? if yes, how was the device removed from the patient? => when a forceps was used, the jaws opened.Did the jaws eventually open?=>yes.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech60c device was returned with no apparent damage and with one gst60g reload loaded on the device.The reload was received fully fired.Upon further inspection, the reload was found to have damaged on the knife channel.The found damage on the reload is consistent when the device is fired over an already existing staple line; when this happens it is possible that the knife may push the staple line and tissue forward shaving the reload deck.The manual override door out of position; the override lever was up which denotes that the knife was manually returned to home position.It should be noted after the manual override system is used the instrument is disabled and cannot be articulated or used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.However, the cut was noted to be jagged due to a damaged knife.One possible cause for this type of damage to the knife is when the device is fired over an already existing staple line, hard object, or thicker tissue than indicated.Repeatedly firing across existing staple lines can also reduce the ability to cut cleanly.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.The event described of damaged closing trigger and difficult to open could not be confirmed as the device opened and closed without any difficulties noted.The closing trigger functioned correctly during functional testing, and no damage was detected.The device event log was reviewed.The log indicates that the device had a high force to fire for one of the clinical firings, indicating that the device was firing over an obstruction, or too thick of tissue.In addition, an event was found that typically indicates that the knife encountered the firing safety lockout mechanism.This mechanism ensures that the knife cannot deploy without an unfired reload installed.It is possible that it occurred due to improper installation of the reload.Always ensure that the retaining cap is in place when a reload is installed into a device.Also, the log indicates that articulation was attempted while the device was closed articulation function is blocked when the device is closed to prevent inadvertent patient injury.When the home button is pressed while the jaws are closed the device will attempt to retract the knife.When any of the other articulation buttons are pressed while the jaws are closed the device will provide haptic feedback indicating that articulation function is blocked.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number a98r9p, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key25350987
Report Number3005075853-2026-04137
Device Sequence Number10748489
Product Code GAG
UDI-Device Identifier10705036034971
UDI-Public(01)10705036034971(17)280930(10)A98R9P
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K213633
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECH60C
Device Lot NumberA98R9P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Initial Report FDA Received Date06/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GST60G
Patient SexUnknown
-
-