• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V+C+S+N+PRO, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device needed service.During the device evaluation, ventec observed that its exhaled tidal volume (vte) levels were low.There were no reports of patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
H6: the device was evaluated by ventec where the reported issue of its exhaled tidal volume (vte) levels being low was confirmed.Ventec replaced the external flow module (efm) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the efm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key25351417
Report Number3013095415-2026-00367
Device Sequence Number21414144
Product Code CBK
UDI-Device Identifier00855573007860
UDI-Public(01)00855573007860(11)201020
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV+C+S+N+PRO, ENGLISH
Device Catalogue NumberPRT-01184-000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2026
Initial Report FDA Received Date06/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-