The investigation determined that lower than expected vitros phenobarbital (phbr) results were obtained from a single american proficiency institute (cap) proficiency survey sample when processed using vitros phbr slide lots: 2520-0118-1165 and 2517-0117-4553 on a vitros xt 7600 integrated system.The assignable cause of the event could not be determined with the information provided.Vitros phbr lot: 1 and lot 2 qc results showed some inaccuracy and imprecision for l2 only, therefore an issue with either of these lots cannot be ruled out as a potential contributor to the event.However, continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros phbr slide lots: 2520-0118-1165 and 2517-0117-4553.An instrument-related performance issue cannot be ruled out as contributor to the event, as no diagnostic precision testing was performed to assess instrument performance at the time of the event.Therefore, an issue with the vitros xt 7600 integrated system cannot be confirmed or ruled out as a contributor to the event.As the customer reported no issues with the other cap samples for vitro phbr, an issue specific to the chm-02 sample cannot be ruled out as a contributor to the events.It was determined that the customer processed the initial vitros phbr sample within the recommended time frame per cap, however no information was provided regarding the sample preparation and storage after receipt and prior to running.Additionally, the vitros phbr repeat was performed outside of the cap stability guidelines of 14 days, therefore an issue isolated to the sample can not be ruled out as a contributor to the lower than expected repeat result.
|