| Catalog Number |
123516A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.(b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that thirty minutes after foley catheter insertion, leakage was observed.Upon inspection, it was found that the catheter bifurcation was damaged.
|
| |
|
Event or Problem Description
|
|
It was reported that thirty minutes after foley catheter insertion, leakage was observed.Upon inspection, it was found that the catheter bifurcation was damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed, cause unknown.The reported failure was able to be reproduced.Based on the attached photo, a tear was observed at the catheter bifurcation area, resulting in water leakage.Further functional testing and comprehensive investigation could not be performed as only photographs were provided for evaluation.Therefore, the exact cause of the issue and the point at which it occurred could not be determined.A potential root cause for this failure could be over dipping caused stripping difficulty, torn funnels and premature deflators and also might be contributed by user-mishandling the product etc.The labeling is found to be adequate.A review of the lot file showed no rejects or rework associated with this issue.For the affected date code, the rejected units were recorded as in process scrap and will be disposed of after inspection.No nonconformities or discrepancies were identified in any of the dhrs.Corrections made to tab(s) f and h.Initial reporter phone: (b)(6).Initial reporter name: (b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|