| Model Number |
6487800 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by distributor's field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the internal leakage test with a message 'system volume too large" failed during pre-use check and replacement of the air gas module solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The air gas module regulates the inspiratory gas flow and a faulty air gas module may affect the delivered volume or gas mixture (o2/air).The device's logs were not provided for further analysis.The defective part has not been returned fot the further investigation, despite request.Our conclusion is that the replaced part was the cause of the failure.Contact person phone number: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed internal leakage test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|