• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NG/NI PEDIATRIC FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NG/NI PEDIATRIC FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 40-8366
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The device history record for lot: 30295817 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.A grouping of small, jagged tears was observed in the tubing, near the connector.The tears are on varying sides of the tubing.The tube was cut laterally, below the base of the enfit connector.Then, cutting as opposite from the tears as possible, the tubing was cut axially, in order to view the inner wall.The inner wall of the tubing exhibits similar tearing and surface abrasions.A root cause has not been established.All information reasonably known as of 01 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported that a hole was noted in nasogastric tubing between the feeding tube and feeding port connection.Event date unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NG/NI PEDIATRIC FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key25354504
Report Number9611594-2026-00380
Device Sequence Number21305220
Product Code KNT
UDI-Device Identifier00350770460369
UDI-Public(01)00350770460369(240)104603602(17)281210(10)30295817
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K831328
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-8366
Device Catalogue Number40-8366
Device Lot Number30295817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2024
Initial Report FDA Received Date06/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age8 MO
Patient SexMale
Patient Weight3 KG
-
-