(b)(4).Date sent: 6/01/2026.D4: batch #101d09.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one vasecr35 reload was received partially fired 1/10th in the lock-out position with the remaining staples inside the reload pockets.It cannot be confirmed why the reload locked out; however, it is possible that the one-piece sled component within the reload may have been moved out of position during the loading process leading to the inadvertent lockout.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position however, the sled stopped at the 1/10th lockout position again.No functional test was performed due to the condition.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 101d09, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|