| Model Number |
PM2172 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The device is included in the field action for assurity, endurity inhomogeneous epoxy mixing issued by abbott on 18 feb 2025 for a subset of devices distributed and implanted outside of the united states.The pacemaker was explanted and replaced prophylactically.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was the prophylactic explant of a device that was subject to the field action for assurity, endurity inhomogeneous epoxy mixing issued by abbott on 18 feb 2025 for a subset of devices distributed and implanted outside of the united states.The device was returned due to advisory with no allegation of a malfunction.Visual inspection of the device and header attachment area did not find any anomalies.Interrogation of the device indicated battery voltage and current within normal range when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).
|
| |
|
Search Alerts/Recalls
|