| Model Number |
V*HOME, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device needed service.Upon evaluation of the device, ventec observed that it was providing inaccurate fio2 (fraction of inspired oxygen) readings.There were no reports of patient involvement associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of it providing inaccurate fio2 (fraction of inspired oxygen) readings was confirmed.Ventec replaced the oa2 board to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined the cause of the reported issue to be the oa2 board.
|
| |
|
Search Alerts/Recalls
|