| Catalog Number |
ASC4830-01 |
| Medical Device Problem Codes |
Degraded (1153); Overheating of Device (1437); Electrical Shorting (2926); Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a surgery, the electrode tip of the super multivac charred and short-circuited, causing intermittent function and severely impacting intraoperative use.It cause overheating of the device.The pieces were removed from the patient.The procedure was resumed, after a non-significant delay, using an equivalent sn back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6 this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the wand tip with significant electrode erosion.One of the electrodes is detached and hanging at the edge of the tip.Additionally, the plastic spacer exhibits burn marks.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the charred electrodes and shorting could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences ((1) hitting the device tip against a hard surface or damage to the tip due to contact with metal objects while activating the wand.2) mechanical displacement of tissue through applied force.3) activating the device above recommended settings for prolonged periods of time or saline ingress).The root cause of the intermittent function and overheating could not be determined since the reported malfunction could not be duplicated during the investigation nor evaluated in the image.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|