U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX MR8; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE Back to Search Results
Model Number MR8-AVA12-R
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the little ball came out of the attachment.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that the little ball came out of the attachment issue was confirmed.The likely cause of the failure is due to internal tube bearing was worn.It was also noted that no abnormalities found.This regulatory report is being submitted due to retrospective review through capa 672570.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX MR8
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25358186
Report Number1625507-2026-01728
Device Sequence Number10653906
Product Code HBE
UDI-Device Identifier00763000063801
UDI-Public(01)00763000063801(11)250508
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AVA12-R
Device Catalogue NumberMR8-AVA12-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2025
Initial Report FDA Received Date06/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-