| Model Number |
MI1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Wound Dehiscence (1154)
|
| Date of Event |
05/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
There was no hearing impression.The user has been explanted.Additionally, there were 2 skin dehiscences and implant infection.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
There was no hearing impression.The user has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found are attributable to the removal surgery.According to the information received from the field the device was explanted due to skin dehiscence above the implant site and infection.Device sterilization records show that the device was subject to a valid sterilization process.Therefore it is unlikely that the device was the source of the reported infection.Additionally, the user never had any hearing benefit most likely due to the previous schwannoma removal.This is a final report.
|
| |
|