| Model Number |
CA1032800 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of the battery failure is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review: the device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Event or Problem Description
|
|
A trilogy 200 was returned to the manufacturer for preventive maintenance.There was no harm or injury reported.There was no evidence the device was in patient use.During the evaluation a service required alarm condition indicating an internal battery was observed in the device's downloaded event log.The internal battery needs to be replaced to address the issue.Additionally, not related to the reportable event, an alarm condition indicating a detachable battery issue was observed.The detachable battery needs to be replaced.The device's rubber feet, handle, base enclosure, humidifier plug, button and keypad need replaced due to missing/physical damage.No components have been replaced due to awaiting parts receipt.A final report is being submitted.If new information becomes available following the completion of the device repair that changes the outcome of the investigation and associated medical device reporting a follow up/supplemental report will be completed.
|
| |
|
Search Alerts/Recalls
|
|
|