The used company specific cartridge was not returned for evaluation.Five samples were randomly selected for evaluation from the returned unopened company specific cartridges for the reported lot.The five unopened company specific cartridges were opened for evaluation.The company specific cartridges were microscopically examined with no damage observed or abnormalities observed.The five company specific cartridges were functionally tested per the ifu (instructions for use).No lens or cartridge damage was observed after the lens deliveries.The company specific cartridges were cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.A photo was provided.The photo showed an implanted single-piece.A scrape mark was observed, which appeared to be on the posterior surface.This damage was similar in appearance to damage caused by a plunger underride.No other determination can be made from the photo.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Qualified associated products were indicated.The root cause for the reported lens damage could not be determined.The used company specific cartridge was not returned for evaluation.A root cause cannot be determined without physical examination of the product.A photo was provided.A scrape mark was observed, which appeared to be on the posterior surface.This damage was similar in appearance to damage caused by a plunger underride.No other determination can be made from the photo.Five samples were randomly selected for evaluation from the returned unopened company specific cartridges for the reported lot.Functional and dye stain testing was conducted with the unopened samples with acceptable results.No lens or cartridge damage was observed after the lens deliveries.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company specific.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company specific iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
|
A physician reported during intraocular lens (iol) implant procedure, a scratch was observed on the posterior side of the lens after implanted.The lens was not removed and remained implanted.The procedure was completed on the same day.There was no patient harm.According to the surgeon, something physically scratched the posterior side of the iol.Additional information has been requested, yet no new information received.
|